HZ3100Electronic weighing instrument calibration
Off State press and hold the [ function ] button and the [ switch ] key to boot

Screen shows "CAH" , At scales Shang not placed any items, by return zero key into correction function, screen displayed "------" into 0 points correction,, screen displayed "XX.XXX Kg" calibration weight value, by peeled key can on last one numerical added a, by cumulative key can overall left moved a bit, digital said should put weight weight, After applies the standard weight corresponding to scales, press zero key, screen shows "------" into the SPAN calibration , later the screen displays "CAH" says correction complete, remove scale weights, press print key, screen shows "EXIT" and press Zero key to exit return to weighing State.

Beijing must supply electronic scales and electronic scale repair service. Maintenance Tel:010-61732587, Helpline:13521224539