Fault code of German sedolis analytical balance
HENGQUASI:2019-11-07 16:06:48

Fault code of German sedolis analytical balance

(BA, GA, BP, GP, IB, PMA, QC, MC1 / Lac / master, MC, LP, LC, AC, La, FB, is Series)

0xgeneral error

Err01 display overflow, output value cannot be displayed.

Due to incorrect menu settings. Reset the menu to factory settings.

Err02 zero position error when adjusting function is activated

The German sedolis analytical balance is not cleared or the weighing plate is unloaded; the zero point is not within the error limit. Check the preload and adjust if necessary.

Err03 zero position error at the end of adjustment

The drift of the sadolis analytical balance is too large and the preheating time is too short. Calibrate after preheating.

Err04 corrects the error many times.

The deviation of each result is too large. Stabilize the German sedolis analytical balance.

Err05 manual calibration weight is not in place.

Check the internal calibration system and send it for repair.

Err06 internal calibration weight failure, or no internal calibration weight found.

Check the internal calibration system for repair.

Err07 is a function or command not allowed in legal metrology.

Consult the sartorius service center. For RC and MC German sedolis analytical balance, air buoyancy correction and marked printing method (721) or (722) are selected.

Err08 tare error or tare memory out of zero range is occupied.

Check the setting of "initial zero range".

Err09 tare error below zero range.

Tare weight cannot be less than zero, clear.

1X tare 1 / tare 2 operation error

Err10 tare 1 is blocked by data under tare 2.

The two tare functions repel each other. Tare 1 is available only after the tare 2 value has been removed.

Err11 tare weight 2 is triggered by a non permissive value.

Negative display value cannot be tare weight. Check the load.

Err12 tare weight 2 > error limit, overflow.

The entered tare weight 2 value is too large, correct it.

Err17 correction error with preload: the required weight value exceeds the upper limit.

Because the load is too heavy to be calibrated internally, reduce the preload or change the setting.

The preload stored by err19 is too heavy.

Reduce preload or change settings.

2X application operation error

The "W" key is pressed before the err20 application is activated, and the application is not activated.

For example: when counting, no reference value is stored.

Err21 parameter is blocked in parameter mode.

Select the menu code (231) and change it to (232) or (233).

Err22 percentage or count weighing active error.

he weighing value is too low or there is no load on the weighing plate.

Err23 application activation error.

Display value is negative or no load on the disk

3x printing error

Err30 prints the key in BPI interface mode.